The materials in the liquid-contact unit are shown below.
FAQDensity/Specific Gravity Meters
Density/Specific Gravity Meters
- Sampling nozzle: Fluorine resin
- Measurement cell: Borosilicate glass
- Built-in syringe: Polypropylene
- Screw plug: Polybutylene terephthalate
- O-ring: Special chemical-resistant fluorine rubber
- Joint connectors: Polyphenylene sulfide
The measurement cell is made of borosilicate glass.
Liquids that dissolve glass such as fluoric acids and highly alkaline samples (sodium hydroxide for example) will seep into the measurement cell. To ensure a long operating life, be careful to clean the cell immediately after measurement.
The polybutylene terephthalate in the screw plug is not resistant to strongly alkaline substances.
If necessary, KEM can change the screw plug to one made of alkali-proof materials.