Drying oven for moisture measurement on chip, particulate or powder samples such as plastic pellets
Maximum set temperature : 300℃
PRODUCT
Evaporator
ADP-611
Features
- The sample boat maneuvers within the heating unit by magnetic control, eliminating hygroscopic permeation, thus allowing an operator to make reliable measurements of trace water.
- Automatic blank heating of sample boat shortens measurement intervals.
- Oven temperature can be easily set by scan mode for optimum evaporating temperature.
- Measurement conditions can be conveniently set on the titrator side.
- Simple structure of the heating tube facilitates cleaning of it.
Specifications
Model name | Evaporator ADP-611 |
---|---|
Heating system | Electrically-conductive transparent heating glass |
Heating range | 50 - 300℃ |
Temperature control | Range: 50 - 300℃ (Minimum settable interva℃) Temperature measurement: K-thermocouple (Precision: ±2℃) |
Display (Temp. / Flow) | 3-digit digital display with LED's on temperature and flow rate |
Heating tube | Pyrex® glass tube: ø30 (O.D.) x 320 (L) [mm] |
Sample boat | Pyrex® glass boat: 68 (L) x 25 (W) x 15 (H) [mm]; Capacity: 16mL |
Carrier gas | Nitrogen gas: Not included in supplied parts Air: Built-in air pump (Optional) |
Drying method on gas | Zeolite tube (100g x 2 units) |
Gas flow rate | 100 - 300 mL/min |
External I/O | Communication with KF titrator via RS232C port (Mini DIN 9-pin) |
Power source | 100-120 VAC or 200-230 VAC, 50/60 Hz |
Power consumption | 150W (Max. 700W) |
Dimension | 312 (W) x 195 (D) x 217 (H) [mm] |
Weight | Approx. 7kg |
Standards
ASTM D1123 Standard Test Methods for Water in Engine Coolant Concentrate by the Karl Fischer Reagent Method
ASTM D1364 Standard Test Method for Water in Volatile Solvents (Karl Fischer Reagent Titration Method)
ASTM D4377 Standard Test Method for Water in Crude Oils by Potentiometric Karl Fischer Titration
ASTM D4672 Standard Test Method for Polyurethane Raw Materials: Determination of Water Content of Polyols
ASTM D6869 Standard Test Method for Coulometric and Volumetric Determination of Moisture in Plastics Using the Karl Fischer Reaction (the Reaction of Iodine with Water)
ASTM E203 Standard Test Method for Water Using Volumetric Karl Fischer Titration
ISO 14897 Plastics -- Polyols for use in the production of polyurethane -- Determination of water content
ISO 7335 Iron ores -- Determination of combined water content -- Karl Fischer titrimetric method
ISO 760 Determination of water -- Karl Fischer method (General method)
JIS C2101 Testing methods of electrical insulating oils
JIS K0068 Test methods for water content of chemical products
JIS K0113 General rules for methods of potentiometric, amperometric, coulometric, and Karl Fischer titrations
JIS K1557-2 Plastics -- Polyols for use in the production of polyurethane -- Part 2: Determination of water content
JIS K2275-2 Crude petroleum and petroleum products -- Determination of water -- Part 2: Potentiometric Karl Fischer titration method
JIS M8211 Iron ores -- Method for determination of combined water content
JIS M8250 Iron ores -- Determination of hygroscopic moisture in analytical samples -- Gravimetric, Karl Fischer titration and mass-loss methods